36.jpg

 

 

Applications
ㆍVenting (filling)
ㆍIsolation (incubators, lyophilizers, EtO sterilizers, fermenters)
ㆍElectronics (gases)

 

사양 - PolyVENT and SteriVENT
Housing Polypropylene
Filter media PTFE (polytetrafluoroethylene)
Pore size 0.2 µm
Vent On inlet
Support system Polypropylene
Sealing Heat-fused
Maximum pressure 2 bar (29 psi) – forward direction
Water break through test 2 bar (29 psi)/15 seconds
Flow direction Bidirectional
Biosafety Materials pass USP Class VI
Sterilization Can be autoclaved at 121°C for 20 minutes (maximum 132°C). Multiple autoclave cycles are possible. However, the responsibility for reuse is with the operator. The device should be protected from cross contamination. An integrity test should be performed after autoclaving. Compatible with EtO sterilization.
Non-pyrogenic LAL total, non-reactive
Filtration area 36 mm capsule: 500 cm2
75 mm capsule: 1000 cm2
150 mm capsule: 2000 cm2
50 mm disc: 16 cm2
25 mm disc: 4 cm2

 

Catalog Number Description
2103 SteriVENT 36, 0.2 µm, inlet 3/8 to 1/2 inch stepped hose barb, outlet 3/8 to 1/2 inch stepped hose barb, 1/pk
2107 SteriVENT 150, 0.2 µm, inlet 3/8 to 1/2 inch stepped hose barb, outlet 3/8 to 1/2 inch stepped hose barb, 1/pk
2108 SteriVENT 150, 0.2 µm, inlet sanitary, outlet sanitary, 1/pk
6713-0425 PolyVENT 4, 50/pk
6713-1075 PolyVENT 1000, 1/pk
6713-1650 PolyVENT 16, 10/pk
6713-1651 PolyVENT 16, 100/pk
6713-5036 PolyVENT 500, 1/pk

LOGIN

SEARCH

MENU NAVIGATION