1. pH measurement |
|
pH range |
-2.000 to 19.999 |
pH resolution (0.1 not in JP) |
0.001/0.01/0.1 |
pH relative accuracy |
+/- 0.001 |
2. mV Measurement |
|
mV range |
-1999 to 1999 |
mV resolution |
0.1 |
mV relative accuracy |
+/- 0.2 |
Relative mV measurement |
Yes |
3. Temperature measurement |
|
Temperature range °C |
-5.0 to 130 |
Temperature resolution °C |
0.1 |
Temperature relative accuracy °C |
+/- 0.1 |
4. Conductivity measurement |
|
Conductivity range, auto ranging |
0.01 ?S/cm to 1000 mS/cm |
Temperature range with blinking display |
80 ? 100 °C |
Conductivity resolution |
2 decimal place |
Conductivity relative accuracy |
+/- 0.5% |
5. Other measurements |
|
TDS range, auto ranging |
0.1 mg/L to 2000g/L |
TDS resolution |
2 decimal place? |
TDS relative accuracy |
+/- 0.5% |
Adjustable solids factor |
Yes |
Resistivity range M?*cm |
0.00 to 2’000 |
Salinity range ppt |
0.00 80.00 |
6. Hardware |
|
Display type? |
Matrix backlit |
Inputs |
BNC, 2mm pin, Mini-DIN, 2 x ATC, power |
Analog outputs |
See S40 |
Communication (RS232 / Ethernet) |
RS 232 / Ethernet? / USB |
Power requirements |
9V DC |
7. Temperature compensation |
|
Manual (Default at 25°C) |
-5.0 to 130.0°C |
Automatic |
-5.0 to 130 |
Simultaneous display |
Yes |
Selectable reference temperature |
20 or 25°C |
Linear correction |
Yes |
Non-linear correction (DIN38404) |
Yes |
8. Calibration functions |
|
Auto buffer recognition / prompting |
Yes |
User defined pH / Conductivity standards |
Yes (9) |
pH calibration |
See S80 |
Conductivity calibration ( 84, 1413, 12.88, sat NaCl) |
1 of 4 points |
Last calibration with time and date |
last 10 /channel |
User defined limits for slope |
Yes (pH only) |
9. Read Format |
|
Manual freeze functions |
Yes |
Reading stable/ready indicator |
Yes |
Auto endpoint |
Yes |
Timed endpoint |
Yes |
Last reading with date and time |
1000 / channel |
Selectable resolution |
Auto ranging |
Display of 25 reading with hi/lo alarms |
Yes |
10. Other functions |
|
Date and time (24 hour clock) continuous |
Yes |
Power loss memory protection |
Yes |
Selectable high/low alarms |
Yes |
Self diagnostics test |
Yes |
Multi-language display 5 selectable |
Yes |
Numeric keypad entry |
Yes |
Alphanumeric input of meter ID, user name, sample ID, sensor ID, serial nr |
Yes |
Data logging capabilities |
1000 /channel |
Barcode reader for sample ID, sensor ID |
Yes |
Selectable RS232 interface settings |
Yes |